首页> 外文会议>International Workshop on Digital Mammography(IWDM 2006) >Volumetric Breast Density Estimation on Mammograms Using Breast Tissue Equivalent Phantoms – An Update
【24h】

Volumetric Breast Density Estimation on Mammograms Using Breast Tissue Equivalent Phantoms – An Update

机译:使用乳房组织当量幻像对乳房X线图的体积乳房密度估计 - 更新

获取原文

摘要

Methods for improving the accuracy of a technique for estimating volumetric breast density are described. A breast tissue-equivalent phantom encompassing a range of thicknesses and compositions of tissue is used to evaluate the sources of error in the technique. The image acquisition parameters that can affect the accuracy of calibration are considered, and sensitivity to these factors is evaluated. The robustness of the technique was tested by obtaining calibration images on 24 mammography machines, at 18 different sites, over a period of 3 years. The ability to use a single calibration on all machines of a given model type was assessed by comparing effective linear attenuation coefficients of fat and fibroglandular tissues, derived from the calibration phantom images obtained from various machines.
机译:描述了提高用于估计体积乳收密度的技术的准确性的方法。包含一系列厚度和组织组成的乳房组织当量模拟物用于评估该技术中的误差源。考虑可能影响校准精度的图像采集参数,评估对这些因素的敏感性。通过在3岁时获得18种不同部位的24种乳房X线摄影机上获得校准图像来测试该技术的稳健性。通过比较从各种机器获得的校准模拟图像的有效线性衰减系数来评估在给定模型类型的所有机器上使用单一校准的能力。