Clinical Evaluation of a Photon-Counting Tomosynthesis Mammography System

机译:光子计数术乳腺X线摄影系统的临床评价

获取原文

摘要

Digital breast tomosynthesis promises solutions to many of the problems currently associated with projection mammography, including elimination of artifactual densities from the superposition of normal tissues and increasing the conspicuity of true lesions that would otherwise be masked by superimposed normal tissue. We have investigated the performance of a novel tomosynthesis system in a clinical setup. The novel system uses 48 photon counting, orientation sensitive, linear detectors which are precisely aligned with the focal spot of the x-ray source. The x-ray source and the digital detectors are scanned in a continuous motion across the patient; each linear detector collecting an image at a distinct angle. The results from an assessment of image quality and the initial clinical trial of this device are presented. Initial results provide anecdotal evidence supporting the superiority of tomosynthesis over projection mammography.
机译:数字乳房Tomosynest承诺对当前与投影乳房X线摄影相关的许多问题的解决方案,包括从正常组织的叠加中消除造成的造成密度,并增加真实病变的阴谋,否则将被叠加的正常组织掩盖。我们已经调查了一种在临床设置中的新型Tomosynthesis系统的表现。新型系统使用48光子计数,取向敏感的线性检测器,其与X射线源的焦点精确对齐。 X射线源和数字探测器在患者穿过患者的连续运动中扫描;每个线性检测器以不同的角度收集图像。提出了一种评估图像质量和该装置的初始临床试验的结果。初始结果提供支持在投影乳房X线摄影上的自成突显的优越性的轶事证据。