Ultrasound Speed in Articular Cartilage under Different Bathing Saline Concentration

机译:不同沐浴盐水浓度下关节软骨的超声速度

获取原文

摘要

Ultrasound techniques have recently been widely used for the characterization of articular cartilage (artC) in vitro and in vivo. The sound speed in artC plays an important role either as an indicator for the artC degeneration or for the calculation of other parameters such as tissue thickness or stiffness. Previous studies on artC have been carried out in bathing saline solutions with different concentrations to investigate swelling behaviors of artC. There is a lack of information in the literature about the effect of bathing saline concentration on the sound speed of artC. In the present study we measured sound speed in artC under different saline concentrations ranging from OM to 2.5M at the room temperature temperature (21 +- 1 deg C). ArtC specimens from bovine patellar models (n=20) were used in these in-vitro studies. Results demonstrated that the sound speed in artC ranged from 1681 +- 50 m/s to 1816 +- 54 m/s when the saline-concentration varied from OM to 2.5M, while the sound speed in saline changed from 1521 +- 3 m/s to 1674 +- 3 m/s. The sound speed linearly (r~2 = 0.99, p <0.001) increased with the increase of the saline concentration at a rate of 55 m/s per mole change. It is concluded that the variations of the bathing saline concentration significantly affect the sound speed in artC and should be well documented in the ultrasonic studies of artC.
机译:最近已经广泛用于在体外和体内表征关节软骨(ARTC)的超声技术。 ARTC中的声速是作为ARTC变性的指示器或用于计算其他参数,例如组织厚度或刚度。以前的ARC研究已经在沐浴盐水溶液中进行,具有不同浓度的浓度,以研究ARTC的肿胀行为。文献中缺乏关于沐浴盐水浓度对ARTC声速的影响的信息。在本研究中,我们在室温温度(21±1℃)的不同盐水浓度下测量ARTC的声速。在这些体外研究中使用来自牛髌骨模型(n = 20)的ARTC标本。结果表明,当盐水浓度从OM变化到2.5米时,ARTC中的声速范围为1681±50米/秒至1816 + - 54米/秒,而盐水中的声速从1521 + - 3米变化/ s至1674 + - 3米/秒。随着每摩尔变化为55米/秒的速率,声速线性(R〜2 = 0.99,P <0.001)增加了55米/秒的速率。得出结论,沐浴盐水浓度的变化显着影响ARTC中的声速,并在ARTC的超声波研究中进行了很好的记录。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号