IN-SITU (ON-SITE) GAS METER PROVING

机译:原位(现场)煤气表证明

获取原文

摘要

Natural gas flow rate measurement errors at field meter stations can result from the installation configuration, the calibration of the meter at conditions other than the actual operating conditions, or the degradation of meter performance over time. The best method for eliminating these or other sources of error is with in-situ (on-site) calibration of the meter. That is, the measurement accuracy of the field meter station should be verified under actual operating conditions by comparing to a master meter or prover. Field provers have been developed for operation at high line pressures and flow rates. [For purposes of this discussion, a high gas flow rate is any flow greater than 3,000 actual cubic feet per hour or (85 m~3/h) at pressures to 1,440 psig (10 MPa).]
机译:现场仪表站的天然气流量测量误差可能是由于安装配置,仪表在实际操作条件以外的条件下校准,或随时间的仪表性能降低。消除这些或其他误差源的最佳方法是仪表的原位(现场)校准。也就是说,通过比较主仪表或先驱,应在实际操作条件下验证现场仪表站的测量精度。在高线压力和流速下开发了现场普通的操作。 [出于本讨论的目的,高气体流速是大于每小时3,000实际立方英尺或(85米〜3 / h)的流量,压力为1,440psig(10MPa)。]