首页> 外文会议>International Symposium on Ceramics in Medicine >Effect of Acetabular Cup Position on the Contact Mechanics of Ceramicon-Ceramic Hip Joint Replacements
【24h】

Effect of Acetabular Cup Position on the Contact Mechanics of Ceramicon-Ceramic Hip Joint Replacements

机译:髋臼杯位置对陶瓷 - 陶瓷髋关节置换接触机械的影响

获取原文

摘要

The importance of the position of an alumina acetabular cup in close contact with an alumina femoral head was examined in this study using the finite element method. The effects of a wide range of inclination angles from 45deg C to 84deg C and anteversion angles from 0deg C to 25deg C were investigated on the predicted contact mechanics at the articulating surfaces. For the range of implantation angles considered in this study, no edge contact was observed at the rim of the acetabular cup, and the effect of the acetabular cup position on the predicted maximum contact pressure was found to be very small (less than 10%). These findings suggest that the edge contact and associated stripe wear observed clinically in ceramic-on-ceramic hip implants should be related to other mechanisms such as micro-separation, rather than to the position of the acetabular cup directly.
机译:使用有限元方法,在该研究中研究了氧化铝髋臼杯的位置与氧化铝股头紧密接触的重要性。在铰接表面的预测接触机制上研究了从45deg C至84deg C的宽范围倾斜角度和从0degc到25degc的反向角度的影响。对于本研究中考虑的植入角的范围,在髋臼杯的边缘处没有观察到边缘触点,并且发现髋臼杯位置对预测的最大接触压力的影响非常小(小于10%) 。这些发现表明,临床上观察到陶瓷上髋关节植入物的边缘接触和相关条纹磨损应与其他机制(如微分离)有关,而不是直接到髋臼杯的位置。