首页> 外文会议>International Symposium on Ceramics in Medicine >In Vitro Dissolution of Amorphous Calcium Phosphate (Acp) Increased the Wear Particle Generation of Plasma-Sprayed HA Coatings
【24h】

In Vitro Dissolution of Amorphous Calcium Phosphate (Acp) Increased the Wear Particle Generation of Plasma-Sprayed HA Coatings

机译:非晶磷酸钙(ACP)的体外溶解增加了血浆喷涂的HA涂层的磨损颗粒产生

获取原文

摘要

Hydroxyapatite (HA) coated total hip joint device has caused concerns of generating wear particles after long-term implantation. We designed a pin-on-disk (POD) test to examine the morphological changes and wear particle generation of plasma sprayed HA coating in vitro. HA coatings were immersed in supplemented α-calf bovine serum for 48 hours. Serum soaked HA coating exhibited significant amount of weight loss due to dissolution of amorphous calcium phosphate (ACP). POD test demonstrated the serum soaked HA coatings presented many micron-sized particles on the surface while the as-received HA coatings maintained good integrity. The generation of wear particles of the serum soaked HA coatings is related to the reduction of the cohesion/adhesion of HA coatings due to the preferential dissolution of ACP.
机译:羟基磷灰石(HA)涂覆的总髋关节装置引起了长期植入后产生磨损颗粒的担忧。我们设计了一个磁盘(POD)测试,检查体外等离子体喷涂HA涂层的形态变化和磨损粒子产生。将HA涂层浸入补充α-Calf牛血清48小时内。由于无定形磷酸钙(ACP)的溶解,血清浸泡的HA涂层表现出显着的重量损失。 POD试验证明血清浸泡的HA涂层在表面上呈现了许多微米尺寸的颗粒,而接收的HA涂层保持良好的完整性。血清浸出的HA涂层的磨损颗粒的产生与由于ACP的优先溶解而减少了HA涂层的凝聚力/粘附。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号