首页> 外文会议>Annual International Conference on Micro Electro Mechanical Systems >Demonstration of Cortical Recording and Reduced Inflammatory Response Using Flexible Polymer Neural Probes
【24h】

Demonstration of Cortical Recording and Reduced Inflammatory Response Using Flexible Polymer Neural Probes

机译:使用柔性聚合物神经探针证明皮质记录和降低炎症反应

获取原文

摘要

We present the fabrication, characterization, use in cortical recording and histological results of a flexible implantable neural probe. The device is microfabricated in polyimide and platinum, allowing for greater flexibility. It incorporates two layers of platinum electrodes, which greatly reduces the size of neural probes and limits the insertion damage. In recording experiments, acute in-vivo measurements were performed in the mouse cortex. Local field potential, single- and multi-neuron activity were simultaneously recorded. We demonstrate using immunohistochemistry techniques reduced inflammation at the implantation site for microfabricated polyimide neural probes. We therefore show that the major advantage of using polymer probes over silicon probes is the reduced damage due to insertion and probe-brain compliance mismatch.
机译:我们介绍了制造,表征,用于皮质记录和柔性可植入神经探针的组织​​学结果。该装置在聚酰亚胺和铂中进行微制酶,允许更大的灵活性。它采用两层铂电极,大大减小了神经探针的尺寸并限制了插入损坏。在记录实验中,在小鼠皮质中进行急性体内测量。同时记录局部场势,单态和多神经元活性。我们用免疫组织化学技术证明在植入部位的植入部位炎症进行微生物的聚酰亚胺神经探针。因此,我们表明,在硅探针上使用聚合物探针的主要优点是由于插入和探针 - 脑顺应不匹配而导致的损坏降低。