首页> 外文会议>Conference on penetrating radiation systems and applications >Measurement of coincidence timing resolution with CdTe detectors
【24h】

Measurement of coincidence timing resolution with CdTe detectors

机译:用CDTE探测器测量符合CDTE的正时分辨率

获取原文

摘要

The coincidence timing resolution is a critical performance parameter to determine if direct gamma-ray detection by semiconductor detector materials such as CdTe or CdZnTe may be suitable for use in positron emission tomography (PET) systems. We report experimental results of the coincidence timing resolution measured with a pair of 2 * 2 * 10 mm~3 CdTe detectors irradiated by 511 keV gamma rays under conditions that allow good timing resolution and good energy resolution to be obtained simultaneously. The measured coincidence time resolutions ranged from 14 ns (350 keV lower energy threshold) to 24 ns (100 keV lower energy threshold). We also found that the coincidence timing resolution was proportional to the signal rise time, and that both of these were proportional to the electron transit time.
机译:重合时序分辨率是临界性能参数,以确定通过诸如CDTE或CDZNTE的半导体检测器材料的直接伽马射线检测是否适用于正电子发射断层扫描(PET)系统。我们报告用一对2×2×10mm〜3个CdTe探测器测量的重合时序分辨率的实验结果,在允许同时获得良好的时序分辨率和良好的能量分辨率的条件下在允许良好的定时分辨率和良好的能量分辨率下进行照射的2×2×10mm〜3个CDTE探测器。测量的巧合时间分辨率范围为14 ns(350kev较低的能量阈值)至24 ns(100keV较低的能量阈值)。我们还发现,巧合时分分辨率与信号上升时间成比例,并且这两者与电子传输时间成比例。