摘要

We report on the characterization of 2 mm thick CdTe diode detector with Schottky contacts to be employed in a novel conceptual design of PET scanner. Results at −8°C with an applied bias voltage of −1000 V/mm show a 1.2% FWHM energy resolution at 511 keV. Coincidence time resolution has been measured by triggering on the preamplifier output signal to improve the timing resolution of the detector. Results at the same bias and temperature conditions show a FWHM of 6 ns with a minimum acceptance energy of 500 keV.These results show that pixelated CdTe Schottky diode is an excellent candidate for the development of next generation nuclear medical imaging devices such as PET, Compton gamma cameras, and especially PET-MRI hybrid systems when used in a magnetic field immune configuration.
机译:我们报告了具有肖特基触点的2 mm厚CdTe二极管探测器的特性,该探测器将用于PET扫描仪的新颖概念设计中。在−8°C且施加的偏置电压为-1000 V / mm时的结果显示,在511 keV时,FWHM的能量分辨率为1.2%。符合时间分辨率已通过触发前置放大器输出信号来测量,以提高检测器的时序分辨率。在相同偏压和温度条件下的结果显示FWHM为6 ns,最小接受能量为500 keV。这些结果表明像素化CdTe肖特基二极管是开发下一代核医学成像设备(如PET,Compton)的极佳候选者伽玛相机,尤其是在磁场免疫配置中使用的PET-MRI混合系统。