The new airborne remote sensing instrument FUBISS measures light intensities from the ultraviolet (200nm) to the near infrared (1,7 um) region of the electromagnetic spectrum with a spectral resolution of up to 2,4 nm in the visible and 10 nm in the near infrared range. A number of parameters have a large influence on the measured spectral data of a discrete spectral light curve and must be corrected. We will present a correction and calibration procedure after a short description of the instrument itself. This is followed by a description of the laboratory measurements and the calculation of the offset and calibration coefficients, finally yielding absolute calibrated spectra with an estimated relative error below 1% for a well-known calibration source.
机译:新的机载遥感仪器FUBISS测量从紫外线(200nm)到电磁谱的近红外(1,7μm)区域的光强度,光谱分辨率高达2,4nm,在可见和10nm近红外线。许多参数对离散光谱光曲线的测量光谱数据具有很大的影响,并且必须校正。在简短描述仪器本身后,我们将提出校正和校准程序。接下来是对实验室测量的描述和偏移和校准系数的计算,最后产生绝对校准光谱,众所周知的校准源的估计相对误差低于1%。