首页> 外文会议>International airborne remote sensing conference amd exhibition;Canadian symposium on remote sensing >CALIBRATION OF FUBISS (FREE UNIVERSITY BERLIN'S INTEGRATED SPECTROGRAPHIC SYSTEM)
【24h】

CALIBRATION OF FUBISS (FREE UNIVERSITY BERLIN'S INTEGRATED SPECTROGRAPHIC SYSTEM)

机译:FUBISS的校准(免费的柏林综合光谱系统)

获取原文

摘要

The new airborne remote sensing instrument FUBISS measures light intensities from the ultraviolet (200nm) to the near infrared (1,7 um) region of the electromagnetic spectrum with a spectral resolution of up to 2,4 nm in the visible and 10 nm in the near infrared range. A number of parameters have a large influence on the measured spectral data of a discrete spectral light curve and must be corrected. We will present a correction and calibration procedure after a short description of the instrument itself. This is followed by a description of the laboratory measurements and the calculation of the offset and calibration coefficients, finally yielding absolute calibrated spectra with an estimated relative error below 1% for a well-known calibration source.
机译:新型机载遥感仪器FUBISS可测量从电磁光谱的紫外线(200nm)到近红外(1.7 um)区域的光强度,其可见光谱分辨率高达2.4 nm,光谱分辨率高达10 nm。近红外范围。许多参数对离散光谱光曲线的测量光谱数据有很大影响,因此必须进行校正。在对仪器本身进行简短描述之后,我们将介绍一个校正和校准程序。接下来是对实验室测量值的描述以及偏移和校准系数的计算,最后得出的是绝对校准的光谱,对于已知的校准源,其估计的相对误差低于1%。