首页> 外文会议>Southern biomedical engineering conference >Biomechanical evaluation of fixation devices for mandibular angle fractures
【24h】

Biomechanical evaluation of fixation devices for mandibular angle fractures

机译:颌骨角骨折固定装置的生物力学评价

获取原文

摘要

The treatment of mandibular angle fractures is plagued with the highest postsurgical complication rate of all mandibular fractures. Multiple plating systems are available. In this study four plating variations were compared in a biomechanical in vitro study. All plating systems were noncompressive Synthes titanium plates. The 3.0 mm reconstruction plate with 6 holes, the 2 mm universal fracture plate with 6 and 8 holes and tension bands and the 6 hole universal fracture plate without tension band were compared. Measurements of gaps at the osteotomy sites showed no significant difference for the plating systems. However the vertical (inferior/superior) displacements were significantly higher for the six-hole plate without tension. The horizontal (bucca/lingual) displacement for the six-hole plate without tension had also highest displacement value. The addition of a tension band was effective to reduce these displacement values. Quantitative computed tomography scans were used to demonstrate a minimal difference in bone quality.
机译:颌骨角骨折的治疗血液中颌骨骨折的最高后期并发症率困扰。可提供多种电镀系统。在本研究中,在生物力学的体外研究中比较了四种电镀变化。所有电镀系统都是非压迫的合成钛板。 3.0 mm重建板,具有6个孔,比较了2mm通用骨折板,具有6和8个孔和张力带的张力和张力带的6个孔通用骨折板。截骨部位的间隙的测量显示电镀系统没有显着差异。然而,对于没有张力的六孔板,垂直(劣势/上级)位移显着高。六孔板没有张力的水平(Bucca /舌)位移也具有最高的位移值。添加张力带是有效的,以减少这些位移值。定量计算断层扫描扫描用于展示骨质质量的最小差异。