Comparison between stationary intensity and lifetime luminescence measurements with fiber optic oxygen sensors containing Ru(4.7Ph2phen)32+ in silicone polymer

机译:硅氧烷聚合物中含有Ru(4.7ph2phen)3 2+的光纤氧传感器静止强度和寿命发光测量的比较

获取原文

摘要

A behavior often registered in intensity based measurements of such sensors are the downcurved STERN-VOLMER-plots, which result in problems of reproducibility and calibration, especially between different sensors of the same type. Another effect is the disagreement between lifetime- and intensity-based STERN-VOLMER-plots, which should be the same in case of dynamic quenching. For a fiber optic system with the $LB@Ru(4.7Ph$- 2$/phen)$-3$/$RB$+2$PLU$/-complex in polydimethylsiloxan on the fiber tip mono- and multi-exponential decay models have been examined. There are no significant differences between monoexponential fitted lifetimes and weighted mean lifetimes from multi-exponential models. STERN-VOLMER-plots from lifetime measurements are in good agreement with intensity based measurements on fresh prepared samples.
机译:通常在基于强度的这种传感器的测量中登记的行为是较次的船尾波动图,这导致重复性和校准的问题,特别是在相同类型的不同传感器之间。另一种效果是在基于寿命和强度的船尾波特之间的分歧,这在动态淬火时应该是相同的。对于带有$ LB @ Ru(4.7ph-$ - 2 $ / phen)的光纤系统 - 3美元/ $ rb $ + 2 $ 5 $ / - 纤维尖端和多指数衰减的聚二甲基硅氧烷中模型已被检查。单烯烃拟合的寿命与来自多指数模型的加权平均寿命之间没有显着差异。来自寿命测量的船尾波特与新鲜制备样品的强度测量非常一致。