Ultrasound Therapy System and Ablation Results Utilizing Miniature Imaging/Therapy Arrays

机译:超声治疗系统和消融结果利用微型成像/治疗阵列

获取原文

摘要

Array-based imaging and therapy systems have several advantages over single element approaches, and it was previously shown that the imaging and therapy functionality may be combined into dual-mode arrays [1]. In this work, minimally invasive, miniature (2.2 mm×50 mm aperture, 3.3 mm diameter) dual-mode linear arrays have been developed into probes with high acoustic power output (100 W, and >120 W/cm{sup}2 at the source), high transmit efficiency (>65% typical), and good imaging performance (3.4 MHz center frequency, 50% fractional bandwidth, >100 mm deep field of view). These therapy / imaging probes have been integrated into a flexible intense ultrasound surgery platform which also includes conventional diagnostic imaging probes. A system architecture has been developed which includes a 64-channel therapy driver with software selection of array aperture and phasing (1/16th wavelength), frequency (0.5 - 8 MHz), drive amplitude (5 W/channel, nominal), mechanical rotational steering (+/-180 degrees), and temporal sequencing/switching of imaging, and therapy modes. The array-based imaging/therapy system has produced encouraging results in preclinical studies of bulk tissue ablation, imaging, and treatment monitoring of liver in vitro and in vivo, examples of which are presented. The system can be applied to ablation of other soft tissue pathologies, e.g. kidney, heart, uterus, brain, GI, etc.
机译:基于阵列的成像和治疗系统通过单个元件方法具有若干优点,并且前面示出了成像和治疗功能可以组合成双模式阵列[1]。在这项工作中,已经开发出具有高声波功率输出(100W,> 120W / cm {SUP} 2的探针的探针,微创,微型(2.2mm×50mm光圈,3.3mm直径)双模线性阵列源头),高发效率(> 65%典型),良好的成像性能(3.4 MHz中心频率,50%分数带宽,> 100 mm深的视野)。这些治疗/成像探针已经集成到柔性强烈的超声手术平台中,该平台还包括常规的诊断成像探针。已经开发了一种系统架构,包括64通道治疗驱动器,具有阵列孔径和相位(1/16波长),频率(0.5 - 8 MHz),驱动幅度(5W /沟道,标称),机械旋转的软件选择转向(+/- 180度),以及成像和治疗模式的时间测序/切换。基于阵列的成像/治疗系统产生了令人鼓舞的导致体外和体内肝脏的散装组织烧蚀,成像和治疗监测的临床前研究,其实例提出。该系统可以应用于烧蚀其他软组织病理,例如软组织病理。肾,心脏,子宫,脑,gi等

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号