首页> 外文会议>Engineering in Medicine and Biology Society, 2004. IEMBS '04. 26th Annual International Conference of the IEEE >Characterization of anatomically compliant linear cardiac lesions created by the CARDIMA® radiofrequency surgical ablation system
【24h】

Characterization of anatomically compliant linear cardiac lesions created by the CARDIMA® radiofrequency surgical ablation system

机译:由CARDIMA®射频外科消融系统产生的符合解剖学的线性心脏病变的特征

获取原文

摘要

This study investigated anatomical cardiac linear lesion characteristics created by the CARDIMA® RF surgical ablation system. We developed an in vitro porcine cardiac model that simulates cardiac tissue during an on-pump open chest procedure. We used the INTELLITEMP® energy management device to deliver RF energy to an 8 electrode, 8 thermocouple deflectable linear surgical probe, to form curvilinear and linear lesions on porcine myocardium. Ablations were at two target temperatures (70° C and 80° C) and durations (60 sec. and 90 sec.). The results show that this system is capable of forming clinically relevant deep and contiguous lesions of desired length.
机译:这项研究调查了由CARDIMA®RF手术消融系统产生的解剖学上的心脏线性病变特征。我们开发了体外猪心脏模型,该模型可在泵上胸腔开放手术过程中模拟心脏组织。我们使用INTELLITEMP®能量管理设备将射频能量传送至8电极,8热电偶可偏转线性手术探针,以在猪心肌上形成曲线和线性病变。消融是在两个目标温度(70°C和80°C)和持续时间(60秒和90秒)下进行的。结果表明,该系统能够形成所需长度的临床相关的深部和连续病变。