首页> 外文会议>Proceedings of the 15th IFAC World Congress: International Federation of Automatic Control >A NEW PACEMAKER CONCEPT FOR RATE-RESPONSIVE PACING BASED ON THE ATRIO-VENTRICULAR CONDUCTION TIME
【24h】

A NEW PACEMAKER CONCEPT FOR RATE-RESPONSIVE PACING BASED ON THE ATRIO-VENTRICULAR CONDUCTION TIME

机译:基于心室传导时间的速率响应式起搏器新概念

获取原文

摘要

The atrio-ventricular conduction time (AVCT), a time interval which can be obtained from an electrocardiogram, was used for rate-responsive pacing in chronotropically incompetent patients. A control loop is always established as AVCT is sensitive to both the level of exertion and the pacing frequency. Based on experiments a set of plant models was derived for each patient. The controller design was governed by the necessity to attenuate the disturbances acting on AVCT which are correlated with the respiratory cycle. In an ongoing experimental study individual controllers are realized for each patient, but with the same closed-loop bandwidth.
机译:房室传导时间(AVCT)是一个可以从心电图获得的时间间隔,用于在时间变通性不佳的患者中进行速率反应性起搏。由于AVCT对运动水平和起搏频率均敏感,因此始终会建立控制回路。根据实验,为每位患者导出了一组植物模型。控制器的设计受制于衰减与呼吸循环相关的作用于AVCT的干扰的必要性。在正在进行的实验研究中,为每个患者实现了单独的控制器,但是具有相同的闭环带宽。