首页> 外文会议>Medical Imaging 2000: Image Processing >Variable-length correlation method for the correction of body motions and heart creep in the SPECT myocardial perfusion imaging
【24h】

Variable-length correlation method for the correction of body motions and heart creep in the SPECT myocardial perfusion imaging

机译:可变长度相关方法在SPECT心肌灌注成像中校正身体运动和心脏蠕动

获取原文

摘要

Abstract: Body motion and heart upward creep are among the most frequent sources of artifacts in the single-photon emission computed tomography (SPECT) of nuclear medicine. This paper provides a new method for automatic correction of such motions. Under the formulation of a variable length correlation, abrupt body motions, gradual body motions, and heart upward creep are corrected in sequential passes. An affine transformation is used to compensate for changes in heart appearance caused by varying view angles in image acquisition. Additionally, a method is proposed for automatic exclusion of non-cardiac organs. The effectiveness of the method has been demonstrated with experimental results. !7
机译:摘要:身体运动和心脏向上蠕动是核医学单光子发射计算机断层扫描(SPECT)中最常见的伪影来源。本文提供了一种自动校正此类运动的新方法。根据可变长度相关性的公式,可以在连续通过中纠正突然的身体运动,逐渐的身体运动和心脏向上蠕动。仿射变换用于补偿由于图像采集中视角变化而引起的心脏外观变化。另外,提出了一种自动排除非心脏器官的方法。实验结果证明了该方法的有效性。 !7