Comparison of AC Power Referenced to Either PJVS or JAWS

机译:参照PJVS或JAWS交流电源的比较

获取原文

摘要

The results of a comparison on active power meter calibrations at the National Institute of Standards and Technology (NIST) using a source of AC power referenced to either a programmable Josephson voltage standard (PJVS) or a Josephson arbitrary waveform synthesizer (JAWS) is presented. The measurements were made at 60 V RMS, 5 A RMS, and 50 Hz at a power factor of 1.0. The estimated uncertainty of the active power applied to a meter under test (MUT) of the two system configurations is within $2 mu mathrm{W}/ext{VA} (mathrm{k}=1)$, and the comparison results demonstrate that the two system configurations agree to within 1 part in 106 for applied voltage, current, and phase. A discussion of the measurement techniques and major sources of error are given.
机译:给出了美国国家标准技术研究院(NIST)使用参考可编程Josephson电压标准(PJVS)或Josephson任意波形合成器(JAWS)的交流电源对有源功率计校准进行比较的结果。在60 V RMS,5 A RMS和50 Hz的功率因数下进行测量。应用于两种系统配置的被测表(MUT)的有功功率的估计不确定度在 $ 2 \ \ mu \ mathrm { W} / \ text {VA} \(\ mathrm {k} = 1)$ ,比较结果表明这两个系统配置在10的1部分中一致 6 适用于施加的电压,电流和相位。讨论了测量技术和主要误差源。