首页> 外文会议>IEEE International Conference on Advanced Intelligent Mechatronics >Design of an ultra thin strain sensor using superelastic nitinol for applications in minimally invasive surgery
【24h】

Design of an ultra thin strain sensor using superelastic nitinol for applications in minimally invasive surgery

机译:使用超弹性镍钛合金的超薄应变传感器设计,用于微创手术

获取原文

摘要

This paper introduces a novel ultra thin strain sensor made of superelastic nitinol wire that is well suited for force sensing applications of surgical instruments. The sensing principle used for the described sensor is the same as that for conventional strain gauges; however, the proposed sensor has significant advantages of thinner size (15 μm diameter), higher gauge factor (3.5), large strain measurement range (up to 4.25%), lower cost and easier installation process. To validate its force sensing capability for minimally invasive surgical instruments, the sensor was mounted on a da Vinci surgical tool to measure the lateral forces acting at the distal end. Experimental results showed that the sensor can accurately measure forces with an RMS error of 32 mN and with a resolution of 55 mN.
机译:本文介绍了一种由超弹性镍钛合金线制成的新型超薄应变传感器,非常适合外科器械的力感测应用。所描述的传感器所使用的传感原理与常规应变仪相同。但是,所提出的传感器具有以下优点:尺寸更小(直径为15μm),标距系数更高(3.5),应变测量范围大(高达4.25%),成本更低且安装过程更容易。为了验证其对微创手术器械的力感测能力,将传感器安装在达芬奇手术工具上,以测量作用在远端的横向力。实验结果表明,该传感器可以以32 mN的RMS误差和55 mN的分辨率准确测量力。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号