摘要

This paper introduces a novel ultra thin strain sensor made of superelastic nitinol wire that is well suited for force sensing applications of surgical instruments. The sensing principle used for the described sensor is the same as that for conventional strain gauges; however, the proposed sensor has significant advantages of thinner size (15 μm diameter), higher gauge factor (3.5), large strain measurement range (up to 4.25%), lower cost and easier installation process. To validate its force sensing capability for minimally invasive surgical instruments, the sensor was mounted on a da Vinci surgical tool to measure the lateral forces acting at the distal end. Experimental results showed that the sensor can accurately measure forces with an RMS error of 32 mN and with a resolution of 55 mN.
机译:本文介绍了一种新型超薄应变传感器,由超弹性镍钛合金电线制成,非常适合于手术器械的力传感应用。用于所描述的传感器的感测原理与传统的应变仪相同;然而,所提出的传感器具有较薄的尺寸(直径为15μm),较高的仪表系数(3.5),大应变测量范围(高达4.25%),更低的成本和更容易安装过程。为了验证其用于微创手术器械的力量感测能力,传感器安装在DA Vinci手术工具上,以测量在远端发挥作用的侧向力。实验结果表明,该传感器可以精确地测量具有32mL的rms误差的力,分辨率为55mN。