Study of fabrication conditions of ATP biosensor based on screen-printed electrode

机译:基于丝网印刷电极的ATP生物传感器的制备条件研究

获取原文

摘要

Adenosine triphosphate (ATP) biosensor has been widely used in biochemistry, molecule biology and food industry in recent years. Meanwhile, screen-printing technology has been applied in electrochemical researches owing to their advanced properties. A dual enzyme ATP biosensor based on co-immobilization of glucose oxidase and hexokinase on screen-printed electrode was developed in this work. The fabrication conditions of platinization and enzyme immobilization, double-layered enzyme electrode structure and magnesium ion concentration during detection were studied. After the optimization, the sensitivity of the ATP biosensor was enhanced up to 20.1 μA mM-1 cm-2. The results will contribute to the fabrication of high sensitivity ATP biosensor.
机译:近年来,三磷酸腺苷(ATP)生物传感器已广泛用于生物化学,分子生物学和食品工业。同时,丝网印刷技术由于其先进的性能已被应用于电化学研究。在这项工作中开发了一种基于将葡萄糖氧化酶和己糖激酶共同固定在丝网印刷电极上的双酶ATP生物传感器。研究了铂化和酶固定化的制备条件,双层酶电极结构和检测过程中镁离子的浓度。优化后,ATP生物传感器的灵敏度提高到20.1μAmM-1 cm-2。结果将有助于高灵敏度ATP生物传感器的制造。