It is well known that iterative reconstruction (IR) of a long object is a challenge in case of circular trajectory. Axial artifacts occur naturally in IR due to data truncation. Because IR uses reprojection, axial coverage of the reconstruction field of view reduces with each iteration. In this work we present a method to extend axial coverage and reduce artifacts in the circular cone-beam scanning geometry. Our approach is based on our recently presented weighted iterative reconstruction algorithm. We use extended FDK reconstruction for FBP seed, expansion of detector rows in Z direction, and a special axial weight designed to reduce artifacts caused by data truncation. Evaluation shows that sharp streaks at the edge of the reconstruction volume are removed and good image quality in the extended region is achieved.
机译:众所周知,在圆形轨迹的情况下,长对象的迭代重建(IR)是挑战。由于数据截断,轴向伪影自然发生IR。因为IR使用重新注入,因此重建的轴向覆盖范围视野和每次迭代都会减少。在这项工作中,我们提出了一种延长轴向覆盖的方法,并减少圆锥梁扫描几何形状中的伪影。我们的方法是基于我们最近呈现的加权迭代重建算法。我们使用扩展FBP种子的FBP改建,在Z方向上扩展探测行,以及特殊的轴向重量,旨在减少由数据截断引起的伪影。评估表明,在重建体积的边缘处的尖锐条纹被移除,并且延伸区域中的图像质量良好。