Linac Leakage Dose Received by Patients Treated Using Non-coplanar Radiotherapy Beams

机译:使用非共面放射疗法束治疗的患者接收的Linac泄漏剂量

获取原文

摘要

As non-coplanar beams are increasingly used to deliver cranial and extra-cranial stereotactic radiotherapy treatments, radiation leakage from the accelerating waveguide, bending magnet and other components of the medical linear accelerator (linac) that are not conventionally brought into proximity to the patient becomes increasingly concerning. In this study, the leakage dose in the patient plane was measured using optically stimulated luminescence dosimeters placed along the treatment couch, at 10 cm intervals. A Varian iX linac was operated in 6 and 10 MV photon mode, with all jaws and multi-leaf collimators closed. Dose measurements were made (a) using a "standard" setup, with couch and gantry at zero degrees and (b) using a worst-case non-coplanar setup, with the couch at 90° and the gantry at 30° (rotated over the couch). Results indicated that the leakage dose in the patient plane was uniformly low (less than 2 cGy/10,000MU) at all measurement positions for both energies, using the standard setup. However, when the gantry and couch were rotated, there was a systematic increase in dose to 4.2 cGy/10,000MU at a point 80 cm from isocentre (below the bending magnet). While these doses are within recommended out-of-field dose limits for static beam treatments, if IMRT/VMAT factors are applied then the leakage dose to the patient from non-coplanar treatments may become unacceptable. Specific checks of out-of-field dose from non-coplanar beam directions are advisable prior to acceptance of new or modified linacs.
机译:由于非共面梁越来越多地用于提供颅骨和颅颅立体定向放射治疗处理,因此来自加速波导,弯曲磁体和医用线性加速器(LINAC)的其他部件的辐射泄漏不常规与患者接近的医用线性加速器(LINAC)变为越来越多地。在该研究中,使用沿着治疗沙发沿10cm的间隔放置的光学刺激的发光剂量计测量患者平面中的泄漏剂量。在6和10 MV光子模式下操作VARIAN IX LINAC,闭合所有钳口和多叶子准直器。使用“标准”设置(a)使用“标准”设置,用零度和(b)的“标准”设置使用最坏情况的非共面积,在90°和30°(旋转)的台式(旋转)沙发)。结果表明,使用标准设置,患者平面中的泄漏剂量在所有测量位置都是均匀的低(小于2cGy / 10,000mu)的漏电剂。然而,当龙门和沙发旋转时,在Isocentre(弯曲磁铁下方)的点80厘米处的点为4.2 cgy / 10,000mu的剂量增加。虽然这些剂量是在推荐的用于静态束处理的场外剂量限制内,但如果应用IMRT / VMAT因子,则从非共面处理的患者泄漏剂量可能变得不可接受。在接受新的或改性的LINAC之前,可以在非共面光束方向上进行特异性检查来自非共面光束方向的外部剂量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号