首页> 外文会议>World Congress on Medical Physics and Biomedical Engineering >Sensitometric analyses of screen-film systems for mammography exams in Brazil
【24h】

Sensitometric analyses of screen-film systems for mammography exams in Brazil

机译:巴西乳房X线摄影考试筛网系统的敏感度分析

获取原文

摘要

A determination of the sensitometric parameters of screen-film systems to evaluate their qualities was performed. The quality control of the automatic film processor was carried out to ensure a high level of efficiency. Based on ISO 9236-3, the following potentials were applied on the X-ray tubes: 25 kV, 28 kV, 30 kV and 35 kV. Four different mammography films from different manufacturers with and without screens were tested for curve shape, speed and average gradient. The results indicated that film 1 exhibited better contrast, film 3 demonstrated the highest energy dependence, and film 4 presented the largest base+fog density. None of the four mammographic films tested achieved satisfactory results in all parameters analyzed. Improvements in the manufacturing process for these films must be completed to avoid losses in the image quality.
机译:进行了筛选筛膜系统的敏感度参数,以评估它们的质量。进行了自动薄膜处理器的质量控制,以确保高效率。基于ISO 9236-3,将以下电位施加在X射线管:25kV,28kV,30 kV和35 kV上。测试来自不同制造商的四种不同的乳房X线摄影膜,用于曲线形状,速度和平均梯度。结果表明,薄膜1表现出更好的对比度,薄膜3证明了最高能量依赖性,薄膜4呈现了最大的基础+雾密度。在分析的所有参数中,测试的四种乳房Xmmography薄膜没有达到令人满意的结果。必须完成这些胶片的制造过程的改进,以避免图像质量的损失。