In vivo liver shear wave motion detection and shear wave speed comparison between fundamental and harmonic imaging

机译:在体内肝脏剪切波动波运动检测和基础和谐波成像之间的剪切波速比较

获取原文

摘要

Quantification of liver elasticity is a major application of shear wave elasticity imaging (SWEI) to non-invasively assess liver fibrosis stages. SWEI measurements can be highly affected by ultrasound image quality. Ultrasound harmonic imaging has shown a significant improvement in ultrasound image quality as well as for SWEI measurements. This has been previously demonstrated in cardiac SWEI. The purpose of this study was to evaluate liver shear wave motion detection and shear wave velocity (SWV) measurements with fundamental and harmonic ultrasound imaging. In a cohort of 14 subjects with no history of liver disease, a 2.5 fold increase in maximum shear wave motion and 0.08 m/s decreases in overall interquartile range and median of SWV, as well as a 14.3% increase in the success rate of SWV measurements was found when using harmonic imaging instead of fundamental imaging.
机译:肝弹性的定量是剪切波弹性成像(SWEI)对非侵入性评估肝纤维化阶段的主要应用。 SWEI测量可以受到超声图像质量的影响。超声谐波成像显示出超声图像质量的显着改善以及SWEI测量。这已在心脏Swei中证明。本研究的目的是评估利用基本和谐波超声成像的肝脏剪切波动检测和剪切波速度(SWV)测量。在没有肝病历史的14个受试者的群组中,最大剪切波动运动的2.5倍增加,总整体夹层范围和SWV中位数下降0.08米/秒,以及SWV成功率增加14.3%使用谐波成像而不是基本成像时发现了测量。