【24h】

Performance of New Dry Sampler for Isocyanates

机译:新型干采样器的异氰酸酯性能

获取原文

摘要

Eleven isocyanates can be simultaneously sampled and analyzed using ASSET EZ4-NCO (dry sampler). Isocyanates form stable DBA-derivatives and the sampler did not need to be processed immediately after spiking or sampling. A comprehensive extraction and analytical method was updated and tested in this study. The use of polypropylene plastic centrifuge tubes of certain brands was found to be acceptable during extraction procedure for shaking and centrifugation steps. Both LC-MS-MS and LC-MS analytical methods were found to be robust and sensitive for the quantitation of isocyanates down to single digit ppb levels.
机译:可以使用ASSET EZ4-NCO(干式采样器​​)同时采样和分析11种异氰酸酯。异氰酸酯形成稳定的DBA衍生物,加标或取样后无需立即处理取样器。在这项研究中,对综合的提取和分析方法进行了更新和测试。发现在摇动和离心步骤的提取过程中使用某些品牌的聚丙烯塑料离心管是可以接受的。已发现LC-MS-MS和LC-MS分析方法对于低至一位数ppb含量的异氰酸酯的定量分析都非常可靠和灵敏。