首页> 外文会议>IEEE Sensors >Measurement of enzyme activity using a plug-based electrochemical microdevice
【24h】

Measurement of enzyme activity using a plug-based electrochemical microdevice

机译:使用基于插头的电化学微型设备测量酶活性

获取原文

摘要

Determination of the activity of an enzyme in a solution of nL order was conducted using an electrochemical device with a plug-based microfluidic processing function. β-galactosidase (β-Gal) and adenosine triphosphatase (ATPase) were used as analyte enzymes. β-Gal is often used to be conjugated with antibodies, whereas the latter plays critical roles in cell functions. The activity of the enzymes was determined by measuring the concentration of one of the final products, p-aminophenol (PAP) or hydrogen peroxide (H2O2) by coulometry. Preparation of plugs of solutions of predetermined volumes and mixing of solutions was carried out using a T-junction on the device. A partially shallow flow channel structure was designed in the sensing area to improve collection efficiency. The observed output charge was correlated well with the amount of PAP or H2O2, which served as an indicator of the enzyme activity.
机译:使用具有基于塞子的微流体处理功能的电化学装置来确定nL级溶液中酶的活性。 β-半乳糖苷酶(β-Gal)和腺苷三磷酸酶(ATPase)被用作分析物酶。 β-Gal通常用于与抗体偶联,而后者在细胞功能中起关键作用。通过库仑法测量最终产物之一的对氨基苯酚(PAP)或过氧化氢(H 2 O 2 )的浓度来确定酶的活性。使用装置上的T型接头制备预定体积的溶液塞并混合溶液。在传感区域设计了部分浅的流道结构,以提高收集效率。观察到的输出电荷与PAP或H 2 O 2 的量很好地相关,PAP或H 2 O 2 可作为酶活性的指标。