首页> 外文会议>2010 IEEE Nuclear Science Symposium : Conference Record >Detection tests of imaging devices based on silicon pixel-array detectors assembled using Tape Automated Bonding and microcable technologies
【24h】

Detection tests of imaging devices based on silicon pixel-array detectors assembled using Tape Automated Bonding and microcable technologies

机译:基于使用磁带自动粘合和微缆技术组装的硅像素阵列检测器的成像设备的检测测试

获取原文

摘要

Detection and spectroscopic tests of three self-triggered silicon detection modules constructed using Tape Automated Bonding and micro-cable technologies and equipped by VATA read-out chips are presented and compared. Results of these tests are demonstrated using hit maps and pulse-height spectra induced by 133Ba, 241Am, and 57Co radioactive sources of X- and gamma rays. We have observed significant improvement in measurement of hit maps in the cases of first and second detector modules. The hit maps of the first detector module have proven that some problems occur in addressing certain VATA chip channels. Whereas, the hit maps of the second detector module is correct. On the other side, spectroscopic resolution given by the Full Width at Half Maximum (FWHM) of the first detection module is below 2 keV which is compatible with measurements on wire bonded modules. The FWHM of the second module is approximately two times worse.
机译:介绍并比较了三个使用磁带自动键合和微电缆技术构建并配有VATA读出芯片的自触发硅检测模块的检测和光谱测试。使用X的 133 Ba, 241 Am和 57 Co放射源诱发的命中图和脉冲高度谱证明了这些测试的结果。 -和伽玛射线。我们已经观察到在第一和第二探测器模块的情况下,命中图的测量有了显着改善。第一个检测器模块的命中图已证明,在解决某些VATA芯片通道时会出现一些问题。而第二检测器模块的命中图是正确的。另一方面,第一个检测模块的半峰全宽(FWHM)给出的光谱分辨率低于2 keV,这与引线键合模块上的测量值兼容。第二个模块的FWHM大约差两倍。

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号