首页> 外文会议>Conference on ultrasonic imaging and signal processing >Quality Control evaluation of 37 liquid crystal displays used in diagnostic services
【24h】

Quality Control evaluation of 37 liquid crystal displays used in diagnostic services

机译:37种液晶显示器的质量控制评估用于诊断服务

获取原文

摘要

The performance of 37 primary class liquid crystal display devices (2, 3 and 5 Mpixel matrix size) used in 9 different diagnostic services in Spain has been determined in terms of 13 quantitative and visual evaluations. The equipment had never been subjected to calibration or to QC tests since commissioning by vendors, between 2 and 18 months before measurements. Tests, using calibrated luminance meters and TG 18 patterns, have evaluated ambient light conditions and other basic performance indicators, namely, display geometric distortion, artefacts, resolution and low-contrast visibility, contrast luminance response compliance to DICOM standard, luminance extreme values and uniformity between pairs of monitors associated to a same workstation. The principal sources of non-compliance are failures to visualize low-contrast test objects (73% of displays), excessive differences with the DICOM contrast response standard (57%), and nonuniform response of monitor pairs (54%). Also, 43% of LCD were found located in places with excessive illumination and presenting specular reflections from faceplates. The analysis of ten 5 Mpixel displays, of possible use in mammography services, indicates similar performance as the rest of monitors, except for the ambient luminance (100% complying with recommendations) and larger non-compliance with the DICOM response standard (80%). No correlation between image quality indicators and monitor hours of operation was found.
机译:37个主要类液晶显示装置(2,3和5mmixel矩阵尺寸)的性能在Spain中的9个不同的诊断服务中已经确定了13个定量和视觉评估。从供应商调试以来,该设备从未经过校准或QC测试,在测量前的2到18个月之间。测试,使用校准亮度仪和TG 18图案,已经评估了环境光条件和其他基本性能指示灯,即显示几何失真,伪影,分辨率和低对比度可见性,对比亮度响应对DICOM标准,亮度极值和均匀性与同一工作站相关联的监视器对之间。不合规性的主要来源是可视化低对比度测试对象(73%的显示器),与DICOM对比度响应标准(57%)的过度差异,以及监测对对的不均匀响应(54%)。此外,43%的LCD被发现位于具有过多照射的地方,并从面板上呈现镜面反射。在乳房X线摄影服务中可能使用的10个M质模显示的分析表明,除了环境亮度(100%遵守建议)和更大的不遵守DICOM响应标准(80%)外,除了其余的监视器的性能。未找到图像质量指示符与监视时间之间的相关性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号