首页> 外文期刊>Medical and Biological Engineering and Computing: Journal of the International Federation for Medical and Biological Engineering >Evaluation of grey-scale standard display function as a calibration tool for diagnostic liquid crystal display monitors using psychophysical analysis.
【24h】

Evaluation of grey-scale standard display function as a calibration tool for diagnostic liquid crystal display monitors using psychophysical analysis.

机译:灰度标准显示的评估可作为使用心理物理分析诊断液晶显示器的校准工具。

获取原文
获取原文并翻译 | 示例
           

摘要

The digital imaging and communications in medicine (DICOM) standard proposed the grey-scale standard display function (GSDF) as a calibration tool for making the gradation characteristic of a radiographic output image consistent. This is designed in such a manner that the contrast visually recognised by observers (called psychophysical contrast) becomes a constant for all digital driving levels. The DICOM standard calls such an ideal characteristic perceptual linearisation. The psychophysical gradient that can express the psychophysical contrast was introduced for the evaluation of the GSDF using a liquid crystal display monitor. Investigations regarding its ability to yield a constant psychophysical contrast, independent of the digital driving level change and under an actual observation environment, such as for clinical radiographic diagnosis in hospital, were carried out. The psychophysical gradients of the GSDF were obtained for two kinds of observation environment: one was a restricted environmentsuch as in a dark room, under steady-state adaptation, using the sinusoidal grading pattern corresponding to the peak frequency of the human eye response. The other was an actual environment reflecting that encountered during clinical diagnosis in a hospital. As a result, the psychophysical gradient under the restricted environment became almost constant and independent of the change in digital driving level, i.e. perceptual linearisation could be satisfied. Furthermore, under the actual observation environment, the psychophysical gradient decreased gradually with the increase in digital driving level, i.e. the perceptual linearisation could not be satisfied. The percentage decrease in the value of the psychophysical gradient at the maximum luminance area was approximately 60% compared with that at the minimum luminance area. Accordingly, the GSDF is unsuitable as a calibration tool for the liquid crystal display monitor, which will be used in actual clinical diagnosis, as it cannot achieve 'perceptual linearisation' under the actual environment. For the purpose of clinical diagnosis, it is necessary to enlarge the physical gradient of GSDF further in the high digital driving level range (which relates to a high luminance area) to give an approximation that is as close to the idealised form as possible.
机译:医学数字成像和通信(DICOM)标准提出了灰度标准显示功能(GSDF)作为校准工具,以使射线照相输出图像的灰度特性保持一致。以这样的方式进行设计:观察者在视觉上识别出的对比度(称为心理物理对比度)对于所有数字驾驶水平而言都是恒定的。 DICOM标准将这种理想的特性感知线性化称为。引入了可以表达心理生理对比的心理生理梯度,用于使用液晶显示器监控GSDF。进行了关于其产生恒定的心理生理对比的能力的研究,该能力与数字驾驶水平变化无关并且在实际观察环境下,例如用于医院的临床放射线诊断。针对两种观察环境获得了GSDF的心理物理梯度:一种是在受限环境下,例如在暗室中,在稳态适应下,使用与人眼响应峰值频率相对应的正弦渐变模式。另一个是反映医院临床诊断期间遇到的实际环境。结果,在受限环境下的心理物理梯度变得几乎恒定,并且与数字驾驶水平的变化无关,即可以满足感知线性化。此外,在实际的观察环境下,随着数字驾驶水平的提高,心理生理梯度逐渐减小,即不能满足感知线性化。与最小亮度区域相比,最大亮度区域的心理生理梯度值的下降百分比约为60%。因此,GSDF不适合用作液晶显示器的校准工具,因为它在实际环境下无法实现“感知线性化”,因此将在实际临床诊断中使用。为了临床诊断的目的,有必要在高数字驱动电平范围(与高亮度区域有关)中进一步扩大GSDF的物理梯度,以使其近似值尽可能接近理想形式。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号