We present the method of the elimination of ECG artifact from the polysomnographic EEG and EGG. The idea of this method is that the ECG-corrupted EEG segment can be detected from ECG R-wave and regarded as a missing segment. After this, we used two interpolations to recover the missing segment. One is the Lagrange polynomial interpolation and the other is the least square error AR interpolation. We also compared the AR-method with the LMS adaptive noise canceling method. Simulations show the AR-method performs better than other methods. To apply to the real EEG and EOG signals practically, we also developed the algorithm to detect whether the artifact level is high or not. If the artifact level is high, then the interpolations are applied.
机译:我们提出了从多导睡眠图脑电图和心电图消除心电图伪影的方法。该方法的思想是,可以从ECG R波中检测出ECG损坏的EEG段,并将其视为丢失的段。此后,我们使用了两次插值法来恢复丢失的片段。一个是拉格朗日多项式插值,另一个是最小二乘误差AR插值。我们还将AR方法与LMS自适应噪声消除方法进行了比较。仿真表明,AR方法的性能优于其他方法。为了实际应用于实际的EEG和EOG信号,我们还开发了一种算法来检测伪影水平是否很高。如果伪影级别较高,则将应用插值。