Influence of doppler-tipped guidewire position in coronary artery on blood flow velocity

机译:冠状动脉中多普勒尖端导丝位置对血流速度的影响

获取原文
获取外文期刊封面目录资料

摘要

Abstract: A pulsatile blood flow model was used to estimate theinfluence of position of Doppler guide wires with 12MHz forward looking ultrasound transducers on theaverage spectral peak velocity. Three 0.014-inch andthree 0.018-inch Doppler guide wires were positioned inplastic tubes ranging from 1.7 mm to 8.0 mm internaldiameter. Blood flow of 50, 100 and 200 ml/min wasadjusted using a roller-pump. The flow velocity waslower by 54% near the wall than in the center of largetubes (diameter 8.0 mm). In tubes of 2.9 mm and 4.2 mmin diameter the maximum variations were 11% and 22.5%for the 0.014-inch guide wire and 7.5% and 20% for the0.018-inch guide wire, respectively. No variance invelocity related to wire position was observed in small(1.7 mm) tubes. The system was not sensitive to angulardisplacement of the guide wire in the range of$POM@30$DGR@. These results demonstrate thatintravascular Doppler ultrasound flowmeter may beaccurately utilized for measurements of blood flowvelocity in small coronary arteries without any need toreposition the guidewire.!10
机译:摘要:采用脉动血流模型估算12MHz前视超声换能器的多普勒导丝位置对平均频谱峰值速度的影响。将三只0.014英寸和三只0.018英寸的多普勒导丝放置在内径为1.7 mm至8.0 mm的塑料管中。使用滚轮泵调节50、100和200 ml / min的血流量。壁附近的流速比大管中心(直径8.0毫米)的流速低54%。在直径为2.9 mm和4.2 mm的管中,0.014英寸导丝的最大变化分别为11%和22.5%,0.018英寸导丝的最大变化分别为7.5%和20%。在小(1.7 mm)管中未观察到与导线位置相关的方差变化率。该系统对在$ POM @ 30 $ DGR @范围内的导丝的角位移不敏感。这些结果表明,血管内多普勒超声流量计可以准确地用于测量小冠状动脉的血流速度,而无需重新放置导丝。10

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号