A Surface Electromyogram Evaluation of the Postural Freedom Effects in Laparoscopic Surgery

机译:腹腔镜手术中姿势自由效应的表面肌电图评估

获取原文

摘要

It has been demonstrated that laparoscopic procedures benefit patients in terms of recovery time, exposure to infections and trauma. Nevertheless, it increases the number of problems for the surgeons, including the frequency and duration of awkward postures for surgeons. The repetition of these movements is considered the main cause for musculoskeletal disorders in surgeons’ upper limbs. The goal of this study is to evaluate the muscular activity and muscular fatigue effect produced by both, a conventional instrument and an instrument provided with the Postural Freedom (PF) feature; which consists in a ball socket articulation that allows a variable handle-to-shaft angle, on a conventional laparoscopic pistol-grip handle. Seventeen participants were evaluated during a static simulation using both instruments. Surface electromyography was used to compare the instruments in terms of muscular activity in each target position and muscular fatigue produced in the muscles trapezius, deltoids, biceps, and flexor carpi radialis. Trapezius and deltoids were the muscles most affected. Entrance and exit targets and targets facing the participants showed the higher muscular activity values. The PF prototype reduced muscular activity in all the muscles and in the majority of the target positions showing a reduction greater than 70% of the activity required by the trapezius and deltoid muscles in comparison to the conventional tool. Muscular fatigue was produced by both instruments but it presented lower frequency values with PF prototype. The results indicated that the use of conventional instruments impacts negatively on muscular activity during laparoscopic procedures, in terms of positions adopted. The PF feature in laparoscopic instrumentation reduced the muscular activity and also decreased the signals of muscular fatigue in the muscles evaluated in comparison with the conventional tool.
机译:已经证明,腹腔镜手术在恢复时间,暴露于感染和创伤方面使患者受益。然而,这增加了外科医生的问题数量,包括外科医生尴尬姿势的频率和持续时间。重复这些动作被认为是外科医生上肢肌肉骨骼疾病的主要原因。这项研究的目的是评估常规仪器和具有姿势自由度(PF)功能的仪器所产生的肌肉活动和肌肉疲劳效果。它包括一个球窝关节,可在传统的腹腔镜手枪式握柄上实现可变的柄与轴角度。使用两种仪器在静态模拟过程中评估了17名参与者。使用表面肌电图来比较器械在每个目标位置的肌肉活动以及斜方肌,三角肌,二头肌和radial屈腕产生的肌肉疲劳。斜方肌和三角肌是受影响最大的肌肉。出入目标和参与者面对的目标显示出较高的肌肉活动值。与传统工具相比,PF原型减少了所有肌肉和大多数目标位置的肌肉活动,显示出斜方肌和三角肌所需的运动减少了70%以上。两种仪器都产生了肌肉疲劳,但PF原型显示出较低的频率值。结果表明,就所采用的姿势而言,常规器械的使用对腹腔镜手术过程中的肌肉活动产生负面影响。与常规工具相比,腹腔镜器械的PF功能减少了肌肉活动,并且还降低了所评估肌肉的肌肉疲劳信号。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号