We propose a new passive pump for high-speed blood collection via ultra-narrow microneedle based on capillary force, in which the hydrophilized microbeads are three dimensionally integrated. The gaps between microbeads act as capillary pumps. Merits of this device are as follows: 1) blood is automatically sucked up when the microneedle reaches to blood vessel, 2) it is much smaller than most conventional syringe pumps. The pumping pressure is inversely proportional to the diameter of each capillary. Flow resistance of the pump decreases with the number of capillaries. Namely, a lot of fine capillaries are effective for increasing the flow rate. A developed pump successfully sucked up human blood via a microneedle (inner dia.; 30μm, length; 1 mm).
机译:我们提出了一种新的被动泵,它基于毛细作用力通过超窄微针进行高速采血,其中亲水化的微珠是三维整合的。微珠之间的间隙充当毛细管泵。该设备的优点如下:1)当微针到达血管时,血液会自动被吸收,2)比大多数传统的注射泵小得多。泵送压力与每个毛细管的直径成反比。泵的流阻随着毛细管数量的增加而降低。即,许多细微的毛细管对于增加流速是有效的。发达的泵通过微针(内径;30μm,长度; 1mm)成功吸收了人体血液。