摘要

Heart rate variability is one of the key parameters for assessing the health status of a subject's cardiovascular system. This paper presents a local model fitting algorithm used for finding single heart beats in photoplethysmogram recordings. The local fit of exponentially decaying cosines of frequencies within the physiological range is used to detect the presence of a heart beat. Using 42 subjects from the CapnoBase database, the average heart rate error was 0.16 BPM and the standard deviation of the absolute estimation error was 0.24 BPM.
机译:心率变异性是评估受试者心血管系统健康状况的关键参数之一。本文提出了一种局部模型拟合算法,用于在光体积描记图记录中查找单个心跳。生理范围内频率的指数衰减余弦的局部拟合用于检测心跳的存在。使用来自CapnoBase数据库的42位受试者,平均心率误差为0.16 BPM,绝对估计误差的标准偏差为0.24 BPM。