Study on Vacuum Leak Calibration Technology Based on Static Cumulative Comparison Method

机译:基于静态累积比较法的真空泄漏校准技术研究

获取原文

摘要

In order to solve the problem of accurate calibration of vacuum leaks with very small leak rate (less than 10-10Pam3/ s), a calibration method based on the static cumulative comparison method is proposed. The method mainly uses a quadrupole mass spectrometer to compare ion current rise rates from standard gas flow meter and vacuum leak during static accumulation, and then the calibration on vacuum leak is achieved. In this paper, the calibration method based on static cumulative comparison is introduced, which includes apparatus principle, calibration process, calibration result processing and so on, and the factors of reducing the measurement uncertainty of calibration result are put forward.
机译:为了解决真空泄漏的精确校准的问题,泄漏率非常小(小于10 -10 p 3 / s)提出了一种基于静态累积比较方法的校准方法。该方法主要使用四极谱质谱仪,从标准气体流量计和真空泄漏在静态累积期间比较离子电流上升速率,然后实现真空泄漏的校准。本文介绍了基于静态累积比较的校准方法,包括设备原理,校准过程,校准结果处理等,提出了降低校准结果的测量不确定度的因素。