首页> 外文期刊>Applied Spectroscopy: Society for Applied Spectroscopy >Comparison of Methods for Transfer of Calibration Models in Near-Infared Spectroscopy: A Case Study Based on Correcting Path Length Differences Using Fiber-Optic Transmittance Probes in In-Line Near-Infrared Spectroscopy
【24h】

Comparison of Methods for Transfer of Calibration Models in Near-Infared Spectroscopy: A Case Study Based on Correcting Path Length Differences Using Fiber-Optic Transmittance Probes in In-Line Near-Infrared Spectroscopy

机译:近红外光谱中校准模型传递方法的比较:基于在线近红外光谱中使用光纤透射探针校正路径长度差异的案例研究

获取原文
获取原文并翻译 | 示例
       

摘要

This article addresses problems related to transfer of calibration models due to variations in distance between the transmittance fiber-optic probes. The data have been generated using a mixture design and measured at five different probe distances. A number of techniques reported in the literature have been compared. These include multiplicative scatter correction (MSC), path length correction (PLC), finite impulse response (FIR), orthogonal signal correction (OSC), piecewise direct standardization (PDS), and robust calibration. The quality of the predictions was expressed in terms of root mean square error of prediction (RMSEP). Robust calibration gave good calibration transfer results, while the other methods did not give acceptable results.
机译:本文讨论了由于透射光纤探头之间的距离变化而导致的校准模型转换相关的问题。数据是使用混合设计生成的,并在五个不同的探针距离处进行了测量。比较了文献中报道的许多技术。其中包括乘法散射校正(MSC),路径长度校正(PLC),有限脉冲响应(FIR),正交信号校正(OSC),分段直接标准化(PDS)和鲁棒性校准。预测的质量以预测的均方根误差(RMSEP)表示。稳健的校准可提供良好的校准传递结果,而其他方法则无法提供可接受的结果。