首页> 外文会议>Annual International Conference of the IEEE Engineering in Medicine and Biology Society >An Noninvasive and Impedance-Ignored Control Strategy of the Ablation Zone in Radiofrequency Ablation Therapy
【24h】

An Noninvasive and Impedance-Ignored Control Strategy of the Ablation Zone in Radiofrequency Ablation Therapy

机译:射频消融疗法消融区的非侵蚀性和阻抗忽略控制策略

获取原文

摘要

This study proposed a control strategy of the ablation margin using the temperature of the probe, without causing additional damage. Compared with other methods, the proposed strategy is real time and impedance-ignored, thus has a better performance in practice. A theoretical model was established to obtain the temperature distribution during the treatment. Several functions were obtained by fitting the results of the simulation model, with which a preset central temperature curve corresponding to a desired ablation zone was determined to regulate the temperature of the control point. Considering the various impedances in practice, a voltage adjustment method according to the error between the preset central temperature and the practical central temperature was proposed to minimize the effect of impedances. At last, the strategy was verified with phantom experiments. The results show that all the temperatures of the control points reached to 50°C at a specific time and kept for a while, which demonstrated the strategy had a good performance within the error range allowed.
机译:本研究提出了使用探针温度的消融余量的控制策略,而不会导致额外的损伤。与其他方法相比,拟议的策略是实时和阻抗忽略,因此在实践中具有更好的性能。建立理论模型以获得治疗期间的温度分布。通过拟合模拟模型的结果获得了几种功能,利用该仿真结果,确定了对应于所需消融区的预设中央温度曲线以调节控制点的温度。考虑到实践中的各种阻抗,提出了根据预设中央温度和实际中心温度之间的误差的电压调节方法,以最小化阻抗的效果。最后,通过幻影实验验证了该策略。结果表明,控制点的所有温度在特定时间达到50°C,并保持一段时间,这表明该策略在允许的误差范围内具有良好的性能。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号