Design of a High Sensitive Non-Faradaic Impedimetric Sensor

机译:设计高敏感的非法达斯阻碍传感器

获取原文

摘要

an electrochemical approach towards identifying antigen - antibody binding interactions is studied by using a non-faradaic impedimetric sensor fabricated on a printed circuit board (PCB) chip. An electrical methodology for detecting protein interactions at ultra-low concentrations (in the femtogram/mL) regime has been demonstrated. Nanoporous alumina with pore diameter of 200nm and pore depth of 250 nm was used as the signal amplifying medium. Cardiac biomarker, brain natriuretic peptide (BNP) was used as the study marker in characterizing the sensor’s sensitivity. A sensitivity of 10 femtogram/mL was determined based on the impedimetric signal response. Sensitivity was determined through Nyquist plot analysis for the non-faradaic interactions of the protein biomolecules. This paper is the first demonstration of clinically relevant limit of detection with the BNP biomarker.
机译:通过使用在印刷电路板(PCB)芯片上制造的非法纳米阻抗传感器研究了鉴定抗原抗体结合相互作用的电化学方法。已经证明了用于检测超低浓度(在牡味/ ML)制的蛋白质相互作用的电气方法。孔径直径为200nm和孔深度为250nm的纳米孔氧化铝作为信号放大介质。心脏生物标志物,脑利钠肽(BNP)用作表征传感器敏感性的研究标志物。基于阻抗信号响应确定10个汇集图/ mL的灵敏度。通过Nyquist Plot分析确定蛋白质生物分子的非法酰基相互作用的敏感性。本文是对BNP生物标志物临床相关的检测极限的第一次演示。