首页> 外文会议>IEEE International New Circuits and Systems Conference >Wireless communication and power transfer system for intracranial neural recording applications
【24h】

Wireless communication and power transfer system for intracranial neural recording applications

机译:用于颅内神经记录应用的无线通信和功率传输系统

获取原文

摘要

This study presents a system which provides a wireless solution to implant powering and bidirectional communication between a neural implant and an external base station. Wireless power transfer (WPT) is realized by using a 4-coil resonant inductive link at 8.6 MHz at a separation distance of 10 mm, which is the average human scalp thickness. Downlink communication at 10 kbps is performed on the same frequency as the WPT by ASK modulation. Uplink communication at 1.8 Mbps which carries digitized neural data is realized by using an integrated transmitter and a discrete receiver in 426–432 MHz band. The transmitter is composed of a cross-coupled symmetrical LC voltage controlled oscillator which is brought to resonance with a loop antenna instead of an integrated inductor. The loop antenna serves also for radiation.
机译:这项研究提出了一种系统,该系统提供了一种无线解决方案,可以为神经植入物和外部基站之间的植入物供电和双向通信提供支持。无线功率传输(WPT)是通过使用8.6 MHz处的4线圈谐振感应链路以10 mm的分隔距离实现的,该距离是人类头皮的平均厚度。通过ASK调制,以与WPT相同的频率执行10 kbps的下行链路通信。通过使用集成的发射器和426-432 MHz频带中的分立接收器,可以实现1.8 Mbps的上行链路通信,该通信承载数字化的神经数据。发射器由交叉耦合的对称LC压控振荡器组成,该振荡器与环形天线而非集成电感器发生谐振。环形天线也可用于辐射。