摘要

The purpose of the research aims to design and construction of infusion device analyzer. This research has adopted the principle of the standard calibration curve relationship between the digital output of the load cell to the volume of the solution at any time, and the pressure acting on the pressure sensor to the digital output of it. The designed and constructed infusion device analyzer was composed of four main parts:1) the control part consisting of micro switches for selecting the operating mode, 2) signal detection part composing of load cells, pressure sensor for measuring the volumetric and flow rate of fluid delivered from infusion devices and occlusion pressure in TV line respectively, 3) the processing part comprising of microcontroller ARM Cortex M3 with C language program and 4) the display part includes touch screen graphic - LCD color and computer interfacing via a USB port to analyze the operation of the infusion devices in the form of Trumpet Curve. The results of functional testing were compared with standard infusion devices analyzer Metron Lagu showed that the average percentage error of flow rate and occlusion pressure were 0.134 percent and 0.132 percent successively. The results of standard calibration of the flow rates by the laboratory of Calibration Services and Environmental Analysis Department, Technology Promotion Association (Thailand - Japan) that has been certified by the international ISO - 17025 shown that the average uncertainty of the 95 percent confidence level and average percentage error were ±1.5 ml/hr and 1.9 percent respectively. The production cost of the prototype is 15,000 baht.
机译:研究目的是设计和建造输液器分析仪。这项研究采用了标准校准曲线关系的原理,该标准校准曲线关系是称重传感器的数字输出在任何时候都与溶液的体积之间的关系,以及作用在压力传感器上的压力与其数字输出之间的关系。设计建造的输液器分析仪由四个主要部分组成:1)控制部分由用于选择操作模式的微动开关组成; 2)组成称重传感器的信号检测部分;用于测量流体的体积和流速的压力传感器分别通过输液设备和电视线中的闭塞压力输送,3)处理部分包括带有C语言程序的微控制器ARM Cortex M3,以及4)显示部分包括触摸屏图形-LCD彩色和通过USB端口的计算机接口以分析小号曲线形式的输液设备的操作。将功能测试的结果与标准输液设备分析仪Metron Lagu进行比较,结果表明,流速和阻塞压力的平均百分比误差分别为0.134%和0.132%。经国际ISO-17025认证的技术促进协会(泰国-日本)校准服务和环境分析部实验室对流量进行标准校准的结果表明,95%置信水平的平均不确定度和平均百分比误差分别为±1.5 ml / hr和1.9%。原型的生产成本为15,000泰铢。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号