Design and construction of infusion device analyzer

机译:输液装置分析仪的设计与施工

获取原文

摘要

The purpose of the research aims to design and construction of infusion device analyzer. This research has adopted the principle of the standard calibration curve relationship between the digital output of the load cell to the volume of the solution at any time, and the pressure acting on the pressure sensor to the digital output of it. The designed and constructed infusion device analyzer was composed of four main parts:1) the control part consisting of micro switches for selecting the operating mode, 2) signal detection part composing of load cells, pressure sensor for measuring the volumetric and flow rate of fluid delivered from infusion devices and occlusion pressure in TV line respectively, 3) the processing part comprising of microcontroller ARM Cortex M3 with C language program and 4) the display part includes touch screen graphic - LCD color and computer interfacing via a USB port to analyze the operation of the infusion devices in the form of Trumpet Curve. The results of functional testing were compared with standard infusion devices analyzer Metron Lagu showed that the average percentage error of flow rate and occlusion pressure were 0.134 percent and 0.132 percent successively. The results of standard calibration of the flow rates by the laboratory of Calibration Services and Environmental Analysis Department, Technology Promotion Association (Thailand - Japan) that has been certified by the international ISO - 17025 shown that the average uncertainty of the 95 percent confidence level and average percentage error were ±1.5 ml/hr and 1.9 percent respectively. The production cost of the prototype is 15,000 baht.
机译:该研究的目的旨在设计和构建输液装置分析仪。该研究采用了在任何时间将负载电池的数字输出到溶液的体积之间的标准校准曲线关系的原理,以及作用在压力传感器上的压力到数字输出。设计和构造的输液装置分析仪由四个主要部分组成:1)控制部件由微开关组成,用于选择操作模式,2)信号检测部件组成的负载电池,压力传感器测量流体的体积和流速的压力传感器从输液装置和电视线中的遮挡压力分别提供,3)处理部分,包括微控制器臂Cortex M3,带有C语言程序和4)显示部分包括通过USB端口的触摸屏图形 - LCD颜色和计算机接口来分析以小号曲线形式的输液装置的操作。与标准输液装置进行比较函数测试的结果分析仪元型LAGU,显示流速和闭塞压力的平均百分比误差连续0.134%和0.132%。由国际ISO - 17025认证的校准服务和环境分析部门,技术促进协会(泰国 - 日本)的实验室标准校准的结果表明,95%的置信水平和95%的平均不确定性平均百分比误差分别为±1.5 ml / hr和1.9%。原型的生产成本为15,000泰铢。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号