The effects of the analysis technique on hepatic iron evaluation using T2#x2217; mapping with magnetic resonance imaging

机译:用T2 *测绘磁共振成像的分析技术对肝脏铁评价的影响

获取原文

摘要

Iron toxicity is the major factor for tissue damage and organ failure in patients with iron overload. T2∗ measurement with magnetic resonance imaging (MRI) has been established as an effective and non-invasive technique for evaluating iron content in the liver. However, different factors related to the adopted image processing criterion affect the resulting measurements. These factors include the exponential fitting model (single exponential, bi-exponential, or exponential + constant) and the size and location of the selected region of interest and whether it includes vasculature, susceptibility artifacts, or is close to the liver boundary. In this study, we investigate the effects of these various factors on T2∗ measurement using calibrated phantoms with different amounts of iron as well as on patients with different degrees of iron overload. The results show various degrees of similarities and differences between different processing techniques, which should be known by the operator for better selection of the image processing parameters and proper interpretation of the resulting measurements.
机译:铁毒性是铁过载患者组织损伤和器官衰竭的主要因素。 T2 *用磁共振成像(MRI)的测量已经建立为评估肝脏中铁含量的有效和非侵入性技术。然而,与所采用的图像处理标准相关的不同因素影响得到的测量。这些因素包括指数拟合模型(单指数,双指数或指数+常数)以及所选择的感兴趣区域的尺寸和位置,以及它是否包括脉管系统,敏感性伪像,或者接近肝脏边界。在这项研究中,我们研究了这些各种因素对T2 *测量的影响,使用校准的幽灵具有不同数量的铁以及不同铁过载程度的患者。结果显示了不同加工技术之间的各种程度的相似性和差异,该方法应该知道,操作员应该知道,以便更好地选择图像处理参数并正确地解释所得到的测量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号