首页> 外文会议>Computing in Cardiology Conference >Wrist-Located Optical Device for Atrial Fibrillation Screening: A Clinical Study on Twenty Patients
【24h】

Wrist-Located Optical Device for Atrial Fibrillation Screening: A Clinical Study on Twenty Patients

机译:心房颤动筛选的腕部光学装置:二十名患者的临床研究

获取原文

摘要

This study aims at evaluating the performances of a wrist-located device to detect atrial fibrillation (AF) based on photoplethysmography (PPG) technology. Twenty patients referred for catheter ablation of cardiac arrhythmias in whom episodes of sinus rhythm (SR) and AF coexisted were screened. Screening included a 12-lead electrocardiogram (ECG) and a PPG device placed at the wrist measuring cardiac pulsatility by means of infrared light. While reference cardiac interbeat (RR) intervals were obtained from the analysis of the ECG signals, RR intervals from the PPG signals were estimated by detecting systolic down-strokes on the optical waveforms. Classification of SR versus AF epochs was obtained via a support vector machine to which features extracted on 10-second windows were provided. Extracted features included mean, standard deviation, minimum, and interquartile range of RR within an epoch. A total number of 2213 epochs (1927 of AF, 286 of SR) were analyzed, providing classification accuracy of 93.85% for the PPG-based classifier and 98.93% for the ECG-based classifier. These preliminary results suggest that a wrist-located PPG-based monitor might be eligible for future screening of AF in large populations.
机译:该研究旨在评估腕带定位装置的性能以检测基于光学质肌造影(PPG)技术的心房颤动(AF)。二十名患者提到了鼻窦节律(SR)和AF共存的窦性心律失常的导管消融患者。筛选包括通过红外光在手腕测量心脏脉冲性处的12引出心电图(ECG)和PPG装置。虽然从ECG信号的分析获得了参考心脏杂交(RR)间隔,但通过检测光波形上的收缩压下划线来估计来自PPG信号的RR间隔。通过支撑载体机获得Sr与AF时期的分类,提供了在10秒窗口中提取的特征。提取的特征包括时代内RR的均值,标准偏差,最小值和界定范围。分析了2213时的2213时期(AF,286的SR),为ECG的分类器提供93.85%的分类精度为93.85%,98.93%。这些初步结果表明,基于腕部的PPG的监视器可能有资格在大人物中筛选AF的筛选。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号