首页> 外文会议>IEEE International Conference on Micro Electro Mechanical Systems >Impedance biosensor for rapid detection of low concentration of E.coli 0157:H7
【24h】

Impedance biosensor for rapid detection of low concentration of E.coli 0157:H7

机译:阻抗生物传感器,用于快速检测E.COLI 0157的低浓度:H7

获取原文

摘要

An Impedance biosensor for rapid detection of low concentration of E.coli O157:H7 was designed, fabricated and tested. The device consists of specially designed focusing and sensing regions embedded in SU8 microchannel. The focusing region, consists of ramped-down vertical and thin-film focusing electrode pair, exerts positive dielectrophoretic forces to concentrate the bacteria towards the center of microchannel while the bulk fluid exits from the outer two channels. The concentrated bacteria will continue to flow toward the sensing region which consists of anti-E.coli antibody coated interdigitated electrode arrays. The biosensor was able to detect concentrations as low as 39 CFU/mL within 2 hours.
机译:设计,制造和测试了阻抗能够快速检测E.COLI O157:H7的耐久性生物传感器。该装置包括嵌入在SU8 MicroChannel中的专门设计的聚焦和传感区域。聚焦区域由斜坡的垂直和薄膜聚焦电极对组成,施加正介电泳力,以将细菌朝向微通道的中心浓缩,而散装流体从外部两个通道出口。浓缩的细菌将继续朝向感测区域流动,该传感区域由抗E.COLI抗体涂覆的互联电极阵列组成。生物传感器能够在2小时内检测低至39 cfu / ml的浓度。