首页> 外文会议>International Workshop on Active-Matrix Flatpanel Displays and Devices >Fabrication of a quartz-crystal microbalance/surface-plasmon-resonance hybrid sensor and taking advantage of their responses for evaluation of thin films and liquids
【24h】

Fabrication of a quartz-crystal microbalance/surface-plasmon-resonance hybrid sensor and taking advantage of their responses for evaluation of thin films and liquids

机译:制造石英晶微观/表面等离子体共振混合传感器,并利用它们对评估薄膜和液体的反应

获取原文

摘要

Hybrid sensors of quartz crystal microbalance (QCM) and surface plasmon resonance (SPR) were prepared, and the film depositions, vapor sorptions to polymer films, and liquid properties were observed. The relationship between the SPR wavelength and QCM frequency shifts depends on both the refractive index and the density of the adsorbed molecule and would give information for identification. Furthermore, the swelling of the film can be monitored using the different laws of nature in the SPR and QCM methods.
机译:制备石英晶体微稳定(QCM)和表面等离子体共振(SPR)的混合传感器,并观察到聚合物膜的膜沉积,蒸气溶液和液体性质。 SPR波长和QCM频率偏移之间的关系取决于吸附分子的折射率和密度,并提供信息的鉴定。此外,可以使用SPR和QCM方法中使用不同的性质定律来监测膜的肿胀。