首页> 外文会议>IEEE International New Circuits and Systems Conference >Exploiting the transient response of a Pirani gauge sensor for beyond atmospheric pressure measurement
【24h】

Exploiting the transient response of a Pirani gauge sensor for beyond atmospheric pressure measurement

机译:利用皮拉尼压力计传感器的瞬态响应进行大气压测量

获取原文

摘要

In this paper, the design of an interface circuit for exploiting the transient response of a Pirani gauge sensor for beyond atmospheric pressure measurement is proposed. Based on the fact that although the thermal conductance of gas saturates for pressure over 100 mbar leading to the saturation of steady response, the transient response of a Pirani gauge sensor is still sensitive owing to the increase of heating capacitance of the gas. An original method is proposed to exploit the transient response. Thanks to the proposed interface circuit, pressures varying from 0.9 bars to 3 bars, which are limited by the model used, can be clearly distinguished by simulation results. Theoretical analysis shows that the measurable pressure range can go beyond that limit.
机译:在本文中,提出了一种用于利用皮拉尼真空计传感器的瞬态响应进行大气压测量之外的接口电路的设计。基于这样的事实:尽管在100 mbar以上的压力下气体的热导率会饱和,从而导致稳定响应的饱和,但由于气体的热容量增加,皮拉尼真空计传感器的瞬态响应仍然很敏感。提出了一种利用瞬态响应的原始方法。由于采用了建议的接口电路,因此,通过仿真结果可以清楚地区分从0.9 bar到3 bar不等的压力,这些压力受所用模型的限制。理论分析表明,可测量的压力范围可以超出该限制。