We report the design, microfabrication and testing of a flexible 256-electrode array for micro-electrocorticography (μECoG) with an electrode pitch of 500 μm. Our μECoG grid is a flexible five-layer parylene MEMS device (two layers of platinum insulated by three layers of parylene) featuring plasma-etched vias and a monolithically integrated parylene cable which is compression-bonded to a fan-out board using anisotropic conductive film (ACF) technology. We have characterized the device by electrochemical impedance spectroscopy in artificial cerebrospinal fluid (aCSF) and recorded acoustic evoked potentials in vivo from the rat primary auditory cortex.
机译:我们报告设计,微型制造和测试的微电皮层照相术(μECoG)的柔性256电极阵列的电极间距为500μm。我们的μECoG网格是一种灵活的五层聚对二甲苯MEMS器件(两层铂由三层聚对二甲苯绝缘),具有等离子蚀刻过孔和单片集成的聚对二甲苯电缆,该电缆通过各向异性导电膜压缩粘合到扇出板上(ACF)技术。我们已经通过在人工脑脊液(aCSF)中的电化学阻抗谱对设备进行了表征,并从大鼠主要听觉皮层体内记录了声诱发电位。