首页> 外文会议>IFIP TC 12 conference on computer and computing technologies in agriculture;CCTA 2010 >Study on the Near Infrared Model Development of Mixed Liquid Samples by the Algorithm of OSC-PLS
【24h】

Study on the Near Infrared Model Development of Mixed Liquid Samples by the Algorithm of OSC-PLS

机译:用OSC-PLS算法研究混合液体样品的近红外模型

获取原文

摘要

The orthogonal signal correction of Wold algorithm was introduced in this thesis, which was combined with the partial least square algorithm. Calibration model of five components of benzene, methylbenzene, chlorobenzene, benzaldehyde and acetophenone in the mixed liquid samples were developed and validated by the method of independent. The result indicated that the calibration models of the components of benzene, methylbenzene, chlorobenzene and benzaldehyde will be developed well after the 1-dimension OSC filter; while a good calibration model of the component of acetophenone will be acquired after the 2-dimension OSC filter.
机译:本文介绍了Wold算法的正交信号校正方法,并将其与偏最小二乘算法相结合。建立了混合液样品中苯,甲基苯,氯苯,苯甲醛和苯乙酮五种成分的校正模型,并采用独立方法进行了验证。结果表明,一维OSC滤光片可以很好地建立苯,甲基苯,氯苯和苯甲醛成分的校正模型。而二维OSC过滤器将获得良好的苯乙酮成分校准模型。