首页> 外文会议>FLOMEKO 2010;International flow measurement conference >Inter-laboratory Comparison Results for Coriolis Mass Flow Meter Calibration Facilities
【24h】

Inter-laboratory Comparison Results for Coriolis Mass Flow Meter Calibration Facilities

机译:科里奥利质量流量计校准设施的实验室间比较结果

获取原文

摘要

Inter-laboratory comparison testing is fundamental to assuring the accuracy of a laboratory's stated measurement uncertainty, and a requirement for accreditation to ISO/IEC 17025:2005. Inter-laboratory comparisons using Micro Motion, Inc. (MMI), Coriolis mass flow meters support the claimed uncertainties of 0.014% (k=2) for the high-accuracy liquid flow calibration facilities, and 0.03% (A=2) for the production flow calibration facilities at Micro Motion, Inc.. This paper describes: multiple calibration methodologies, international traceability to National Metrology Institutes, test results and future design requirements of an inter-laboratory test program required to support multiple, accredited flow calibration centers around the globe.
机译:实验室之间的比较测试是确保实验室陈述的测量不确定度准确性的基础,也是对ISO / IEC 17025:2005认证的要求。使用高准公司(MMI)的科里奥利质量流量计进行的实验室间比较支持高精度液体流量校准设施的不确定度为0.014%(k = 2),而高精度液体校准设施的不确定度为0.03%(A = 2)。高准公司生产流量校准设施。本文描述:多种校准方法,对国家计量学会的国际可追溯性,实验室间测试计划的测试结果和未来设计要求,以支持围绕该认证中心的多个认可的流量校准中心。地球。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号